Packaging Technologies & Inspection (PTI)

VeriPac

VeriPac 225/4

Specifications

Application
Non-destructive leak detection for packages with defect profile typically > 25 microns
Package Type
  • Filled & sealed pouches
  • Flexible & semi-rigid packaging
Industries: Markets Served
Medical Device, Pharmaceutical, Food
Package Materials& Combinations
Tyvek®, Paper, Foil, Film, Aluminum, Plastic & Laminated Materials
On-line or Off-line Testing
Semi-automatic online or offline leak testing applications
Test Components
Absolute Vacuum Transducer
Technology
Vacuum Decay
Operator Interface
Touch Screen (Bilingual English and Spanish)
Test Sensitivity
Manual or Automatic
Test Result Data
Pass/Fail Result - Absolute vacuum measured in mbar/sec., Differential vacuum measured in Pa/sec.
Statistics & Test Results Trends
Yes, including result trends for each test station
Data Collection
  • View test result data on touch screen
  • Electronic data collection (unlimited test data)
Test Chamber
Four Stations - product loading, product testing, discharge and reject
ASTM Test Method
ASTM F2338
FDA Standard
#014; Item 168
Package Defect Detected
Measures whole seal integrity and leak tests entire package
Test Frame Construction
Stainless Steel
Power
100-240 VAC; 50/60 Cycles
Main Air
Maximum 90 psi
Options
  • Microcalibrator flowmeter
  • Validation qualification package (IQ/OQ/PQ)
Warranty
Standard One Year Parts and Labor
 

*ASTM Test Method F2338 - www.astm.org based on VeriPac PTI-225 and PTI-325 leak testers.

Test results may vary according to application and package specifications.