VeriPac
VeriPac 225/4
Specifications
- Application
- Non-destructive leak detection for packages with defect profile typically > 25 microns
- Package Type
-
- Filled & sealed pouches
- Flexible & semi-rigid packaging
- Industries: Markets Served
- Medical Device, Pharmaceutical, Food
- Package Materials& Combinations
- Tyvek®, Paper, Foil, Film, Aluminum, Plastic & Laminated Materials
- On-line or Off-line Testing
- Semi-automatic online or offline leak testing applications
- Test Components
- Absolute Vacuum Transducer
- Technology
- Vacuum Decay
- Operator Interface
- Touch Screen (Bilingual English and Spanish)
- Test Sensitivity
- Manual or Automatic
- Test Result Data
- Pass/Fail Result - Absolute vacuum measured in mbar/sec., Differential vacuum measured in Pa/sec.
- Statistics & Test Results Trends
- Yes, including result trends for each test station
- Data Collection
-
- View test result data on touch screen
- Electronic data collection (unlimited test data)
- Test Chamber
- Four Stations - product loading, product testing, discharge and reject
- ASTM Test Method
- ASTM F2338
- FDA Standard
- #014; Item 168
- Package Defect Detected
- Measures whole seal integrity and leak tests entire package
- Test Frame Construction
- Stainless Steel
- Power
- 100-240 VAC; 50/60 Cycles
- Main Air
- Maximum 90 psi
- Options
-
- Microcalibrator flowmeter
- Validation qualification package (IQ/OQ/PQ)
- Warranty
- Standard One Year Parts and Labor
*ASTM Test Method F2338 - www.astm.org based on VeriPac PTI-225 and PTI-325 leak testers.