VeriPac
Specifications
- Application
- Non destructive leak detection of multi-cavity blister packaging
With defect profile typically >5 microns - Package Type
- Blister packaging: semi-rigid, flexible
Standard configuration: Up to a maximum of 30 cavities
Test multiple packages in one test cycle - Industries: Markets Served
- Pharmaceutical, nutriceutical
- Test Cycle
- Rapid test time
- Blister Changeover
- Fast changeover from one blister format to another
- Package Materials
- Paper, foil, film, plastic – any combination
Printed or non-printed surface, glossy or flat finish - Online or Offline Testing
- Offline lab instrument
- Test Components
- Absolute vacuum transducer
Special camera - Technology
- Vacuum decay and special imaging technology (patents pending)
- Operator Interface
- Touch Screen Windows Based PC
- Test Parameters
- Pre-set parameters for each blister tooling with auto recognition
- Test Result Data
- Pass/Fail result – vacuum decay results measured in Mbar/sec. Identifies defective blister cavities in indirect image
- Statistics & Test Diagnostics
- Yes
- Data Collection
- PC based data log collection – network ready
Print result data via external printer (option) - Test Chamber Tooling
- Drawer mechanism accommodates tooling inserts for each type/size blister package
- Package Defect Detected
- Measures seal and package integrity of entire blister package and identifies defective blister cavities
- Other Notable Features
- Based on ASTM vacuum decay leak test method* with addition of vision technology to ID defective blister cavity
- Test Frame Construction
- Stainless Steel
- Dimensions
- 25” W – 25” D – 36” H
- Weight
- 275 lbs.
- Power
- 100-240 VAC; 50/60 Cycles
- Main Air
- Maximum 90 psi
- Options
- Microcalibrator flowmeter
Validation package (IQ/OQ/PQ) - Warranty
- Standard One Year Parts and Labor
*ASTM Test Method F2338-05 - www.astm.org based on VeriPac PTI-225 and PTI-325 leak testers.