Packaging Technologies & Inspection (PTI)

VeriPac 325

Application
Non-destructive leak detection for packages with defect profile typically >10 microns
Package Type
  • Filled & sealed pouches containing dry product
  • Rigid containers
  • Lidded flexible & semi-rigid packaging made of non-porous materials
Package Materials& Combinations
Film, Aluminum, Plastic & Laminated Materials
Test Configuration
  • Offline laboratory
  • Production line applications
Test System
Dual Transducers
Vacuum Source
Internal vacuum
Technology
Differential Pressure & Vacuum Decay
Operator Interface
Touch Screen
Test Parameter Storage
Up to 30 packages
Test Sensitivity
1.5 ccm (approximate hole size 10 microns)
Test Result/Resolution
Pass/Fail Result - measures differential pressure & vacuum decay in Pascal/sec
Test Result Trends
Yes
Data Collection
View on touch screen and electronic
Test Chamber
Manual or semi-automatic
ASTM Test Method
ASTM F2338-09*
Test Frame Construction
Stainless Steel
Tester Dimensions
12” W – 18.5” D – 10” H
Weight
28 lbs
Power
100-240 VAC; 50/60 Cycles
Air
90 psi
Options
  • Validation Qualification Package (IQ/OQ/PQ)
  • Microcalibrator Flowmeter
 

*ASTM Test Method F2338-09 - www.astm.org based on VeriPac leak testers. Test results may vary according to application and package specifications.

Test results may vary according to application and package specifications.