VeriPac 325/D
Specifications
- Application
- Non-destructive leak detection for dry filled packages with defect profile typically >10 microns
- Package Type
-
- Pouches and flexible packaging made of non-porous materials
- Poly peelable chevron pouches
- Stick packs and sachets
- Package Materials & Combinations
- Non-porous materials: foil, plastic, poly, film, Aluminum
- Test Configuration
-
- Offline laboratory
- Production line applications
- Test System
- Dual Transducer Technology
- Vacuum Source
- Internal vacuum
- Test Method*
- Differential Vacuum Decay
- Test Vacuum Range
- Down to -750 mBar Vacuum (22 inHg)
- Operator Interface
- 6" Color Touch Screen
- Test Parameter Storage
- Up to 30 packages
- Test Sensitivity
- 1 ccm (approximate hole size 10 microns)
- Test Results/Resolution
- Pass/Fail Result in mBar and Pascal/sec
- Security Password
- No
- Remote Internet Access
- No
- Data Collection
- View on touch screen and electronic
- Test Chamber
- Test drawer with flexible membrane: Large drawer (24.5" x 14.5") or Small drawer (8.5" x 14.5")
- ASTM Test Method
- ASTM F2338-09**
- Test Instrument Enclosure
- Stainless Steel
- Tester Dimensions/Weight
- 12" W - 18.5" D - 10" H / 28 lbs.
- Test Drawer Dimensions/Weight
- Small Drawer: 17" W x 21½" D - 133 lbs. / Large Drawer: 34" W x 25½" D - 227 lbs.
- Power
- 100-240 VAC; 50/60 Cycles
- Air
- 90 psi
- Options
- Microcalibrator flowmeter
Validation qualification package (IQ/OQ/PQ)
* U.S. Patent Pending.
** ASTM Test Method F2338-09 - www.astm.org based on VeriPac leak testers.
Test results may vary according to application and package specifications.