VeriPac 325/LV
Specifications
- Application
- Micro Leak Detection
Container Closure Integrity Testing - Package Type
- Empty & pre-filled syringes
Liquid filled & lyophilized vials
Filled & sealed bottles, cups, FFS bottles, ampoules
Non-porous pouches
BPC (Bulk Pharmaceutical Chemical) containers
API (Active Pharmaceutical Ingredient ) containers
BFS containers - Industries: Markets Served
- Pharmaceutical, nutritional and others
- Online or Offline Testing
- Offline lab instrument or unline leak testing applicactions
- Test Components
- Absolute vacuum & differential vacuum transducers
- Technology
- Differential pressure & vacuum decay
- Operator Interface
- Touch Screen
- Test Parameter Storage
- Up to 30 packages
- Test Sensitivity
- Down to 0.01 cc/min. (1 - 1.5 micron)
- Test Results
- Pass/Fail Result - measures differential pressure
& vacuum decay in Pascale units. - Data Collection
- Collects test data for view on touch screen
and electronic data collection. - Test Chamber Tooling
- Manual or automatic
- ASTM Test Method Approved
- ASTM F2338-07*
- Package Defect Detected
- Measures whole seal integrity of container or package and quality of entire package
Measures and verifies container closure system integrity - Test Frame Construction
- Stainless Steel
- Dimensions
- 20" W – 216" D – 39.5" H
- Weight
- 28 lbs.
- Power
- 100-240 VAC; 50/60 Cycles
- Main Air
- Maximum 90 psi
- Options
- Microcalibrator flowmeter
Validation qualification package (IQ/OQ/PQ)
*ASTM Test Method F2338-07 - www.astm.org based on VeriPac PTI-225 and PTI-325 leak testers.