Packaging Technologies & Inspection (PTI)

VeriPac 325/LV

 

Specifications

Application
Micro Leak Detection
Container Closure Integrity Testing
Package Type
Empty & pre-filled syringes
Liquid filled & lyophilized vials
Filled & sealed bottles, cups, FFS bottles, ampoules
Non-porous pouches
BPC (Bulk Pharmaceutical Chemical) containers
API (Active Pharmaceutical Ingredient ) containers
BFS containers
Industries: Markets Served
Pharmaceutical, nutritional and others
Online or Offline Testing
Offline lab instrument or unline leak testing applicactions
Test Components
Absolute vacuum & differential vacuum transducers
Technology
Differential pressure & vacuum decay
Operator Interface
Touch Screen
Test Parameter Storage
Up to 30 packages
Test Sensitivity
Down to 0.01 cc/min. (1 - 1.5 micron)
Test Results
Pass/Fail Result - measures differential pressure
& vacuum decay in Pascale units.
Data Collection
Collects test data for view on touch screen
and electronic data collection.
Test Chamber Tooling
Manual or automatic
ASTM Test Method Approved
ASTM F2338-07*
Package Defect Detected
Measures whole seal integrity of container or package and quality of entire package
Measures and verifies container closure system integrity
Test Frame Construction
Stainless Steel
Dimensions
20" W – 216" D – 39.5" H
Weight
28 lbs.
Power
100-240 VAC; 50/60 Cycles
Main Air
Maximum 90 psi
Options
Microcalibrator flowmeter
Validation qualification package (IQ/OQ/PQ)
 
 

*ASTM Test Method F2338-07 - www.astm.org based on VeriPac PTI-225 and PTI-325 leak testers.

Test results may vary according to application and package specifications.