VeriPac 410
Specifications
- Application
- Non-destructive leak detection of blister packs, sachets and pouches with low headspace
with defect profile typically >15 microns - Package Type
-
- Blister packs
- Sachets
- Pouches with low headspace
- Package Materials
- Non-porous materials: foil, plastic, poly, film, Aluminum, paper
- Offline or Online
- Offline lab instrument
- Test Method*
-
- Absolute vacuum transducer
- Pressure plate
- Technology
- Vacuum decay and differential force measurement
- Operator Interface
- 10" color touch screen
- Test Parameter Storage
- Up to 50 packages
- Test Result Data
- Pass/Fail Result in mBar units
- Test Sensitivity
- 2 ccm (approximate hole size 15 microns)
- Security Password
- Yes
- Remote Internet Access
- Yes
- Data Collection
- View on touch screen and electronic data log collection
- Test Chamber
- Test drawer configuration
- ASTM Test Method
- ASTM F2338-09**
- Test Instrument Enclosure
- Stainless steel
- Test Dimensions/Weight
- 14.5" W x 22" D x 12" H/35 lbs.
- Test Drawer Footprint
- 17" W x 21" D
- Power
- 100-240 VAC;50/60 cycles
- Air
- 90 psi
- Options
- Microcalibrator flowmeter
Validation package (IQ/OQ/PQ)
* U.S. Patent Pending
**ASTM Test Method F2338-09 - www.astm.org based on VeriPac leak testers.
Test results may vary according to application and package specifications.