Packaging Technologies & Inspection (PTI)

VeriPac 410

 

Specifications

Application
Non-destructive leak detection of blister packs, sachets and pouches with low headspace
with defect profile typically >15 microns
Package Type
  • Blister packs
  • Sachets
  • Pouches with low headspace
Package Materials
Non-porous materials: foil, plastic, poly, film, Aluminum, paper
Offline or Online
Offline lab instrument
Test Method*
  • Absolute vacuum transducer
  • Pressure plate
Technology
Vacuum decay and differential force measurement
Operator Interface
10" color touch screen
Test Parameter Storage
Up to 50 packages
Test Result Data
Pass/Fail Result in mBar units
Test Sensitivity
2 ccm (approximate hole size 15 microns)
Security Password
Yes
Remote Internet Access
Yes
Data Collection
View on touch screen and electronic data log collection
Test Chamber
Test drawer configuration
ASTM Test Method
ASTM F2338-09**
Test Instrument Enclosure
Stainless steel
Test Dimensions/Weight
14.5" W x 22" D x 12" H/35 lbs.
Test Drawer Footprint
17" W x 21" D
Power
100-240 VAC;50/60 cycles
Air
90 psi
Options
Microcalibrator flowmeter
Validation package (IQ/OQ/PQ)
 
 

* U.S. Patent Pending
**ASTM Test Method F2338-09 - www.astm.org based on VeriPac leak testers.

Test results may vary according to application and package specifications.