VeriPac 415
Specifications
- Application
- Non-destructive leak detection for dry filled packages with defect profile typically >15 microns
- Package Type
-
- Filled & sealed pouches, cups, trays
- Empty trays
- Tyvek® lidded blisters
- Pouches
- Package Materials & Combinations
- Tyvek®, Paper, Foil, Film, Aluminum, Plastic & Laminated Materials
- Test Configuration
-
- Offline laboratory
- Production line applications
- Test System
- Single Transducer Technology
- Vacuum Source
- Integrated vacuum pump
- Test Method*
- Vacuum Decay
- Test Vacuum Range
- Down to 250 mBar Absolute Vacuum (22 inHg)
- Operator Interface
- 10" Color Touch Screen
- Test Parameter Storage
- Up to 50 packages
- Test Sensitivity
- 2 ccm (Approximate hole size 15 microns)
- Test Results/Resolution
- Pass/Fail Result in mBar units
- Security Password
- Yes
- Remote Internet Access
- Yes
- Data Collection
- View on touch screen and electronic
- Printer Option
- Yes
- Test Chamber
- Manual or semi-automatic
- ASTM Test Method
- ASTM F2338-09**
- Test Instrument Enclosure
- Stainless Steel
- Tester Dimensions
- 14.5” W – 22” D – 12” H
- Weight
- 35 lbs
- Power
- 100-240 VAC; 50/60 Cycles
- Air
- Maximum 90 psi
- Options
-
- Validation Qualification Package (IQ/OQ/PQ)
- Microcalibrator Flowmeter
* U.S. Patent Pending.
** ASTM Test Method F2338-09 - www.astm.org based on VeriPac leak testers.
Test results may vary according to application and package specifications.