Packaging Technologies & Inspection (PTI)

VeriPac 425

Specifications

Application
Non-destructive leak detection for dry filled packages with defect profile typically >5 microns
Package Type
  • Filled & sealed bottles, cups, trays
  • Empty trays & cups
  • Pouches
Package Materials & Combinations
Foil, Film, Aluminum, Plastic & Laminated Materials
Test Configuration
  • Offline laboratory
  • Production line applications
Test System
Dual Transducer Technology
Vacuum Source
Integrated vacuum pump
Test Method*
Differential Vacuum Decay
Test Vacuum Range
Down to 250 mBar Absolute Vacuum (22 inHg)
Operator Interface
10" Color Touch Screen
Test Parameter Storage
Up to 50 packages
Test Sensitivity
0.2 ccm (Approximate hole size 5 microns)
Test Results/Resolution
Pass/Fail Result in mBar and Pascal units
Security Password
Yes
Remote Internet Access
Yes
Data Collection
View on touch screen and electronic
Printer Option
Yes
Test Chamber
Manual or semi-automatic
ASTM Test Method
ASTM F2338-09**
Test Instrument Enclosure
Stainless Steel
Tester Dimensions
14.5” W – 22” D – 12” H
Weight
35 lbs
Power
100-240 VAC; 50/60 Cycles
Air
Maximum 90 psi
Options
  • Validation Qualification Package (IQ/OQ/PQ)
  • Microcalibrator Flowmeter
 

* U.S. Patent Pending.
** ASTM Test Method F2338-09 - www.astm.org based on VeriPac leak testers.

Test results may vary according to application and package specifications.