Application
Non-destructive leak detection for dry filled packages with defect profile typically >5 microns
Package Type
  • Pouches and flexible packaging made of non-porous materials
  • Poly peelable chevron pouches
  • Stick packs and sachets
Package Materials & Combinations
Non-porous materials: foil, plastic, poly, film, Aluminum
Test Configuration
  • Offline laboratory
  • Production line applications
Test System
Dual Transducer Technology
Vacuum Source
Integrated vacuum pump
Test Method*
Differential Vacuum Decay
Test Vacuum Range
Down to 250 mBar Absolute Vacuum (22 inHg)
Operator Interface
10" Color Touch Screen
Test Parameter Storage
Up to 50 packages
Test Sensitivity
0.2 ccm (approximate hole size 5 microns)
Test Results/Resolution
Pass/Fail Result in mBar and Pascal units
Security Password
Yes
Remote Internet Access
Yes
Data Collection
View on touch screen and electronic
Test Chamber
Test drawer with flexible membrane: Large drawer (24.5" x 14.5") or Small drawer (8.5" x 14.5")
ASTM Test Method
ASTM F2338-09**
Test Instrument Enclosure
Stainless Steel
Tester Dimensions/Weight
14.5" H – 22" D x 12" H / 35 lbs.
Test Drawer Dimensions/Weight
Small Drawer: 17" W x 21½" D - 133 lbs. / Large Drawer: 34" W x 25½" D - 227 lbs.
Power
100-240 VAC; 50/60 Cycles
Air
90 psi
Options
Microcalibrator flowmeter
Validation qualification package (IQ/OQ/PQ)
 
 

* U.S. Patent Pending.
** ASTM Test Method F2338-09 - www.astm.org based on VeriPac leak testers.