HOME / PACKAGE INTEGRITY / Container / Veripac-310

VeriPac 310 Package Integrity Test System

Non-destructive alternative to water bath leak test method

SPECIFICATIONS

Application Non-destructive leak detection for dry filled packages with defect profile typically >20 microns
Package Type
  • Filled & sealed pouches containing dry product
  • Rigid containers, cups, trays
  • Lidded flexible & semi-rigid packaging made of non-porous materials
Package Materials & Combinations Film, Aluminum, Plastic & Laminated Materials
Test Configuration
  • Offline laboratory
  • Production line applications
Test System Absolute vacuum transducer with automated pressure regulator
Technology Vacuum Decay
Recognized Test Method ASTM F2338-09, referenced in USP <1207>
Operator Interface Touch Screen
Test Parameter Storage Up to 20 packages
Test Sensitivity 3.4 ccm (approximate hole size 20 microns)
Test Results/Resolution Pass/Fail Result in mBar
Test Result Trends Yes
Data Collection
  • Test results on touch screen up to 300 tests.
  • Unlimited test data via electronic data storage
  • Data transfer via Ethernet cable or SD card
Test Chamber Manual or semi-automatic
Test Instrument Enclosure Stainless Steel
Tester Dimensions 12” W – 18.5” D – 10” H
Weight 30 lbs
Power 100-240 VAC; 50/60 Cycles
Air 90 psi
Options
  • Validation Qualification Package (IQ/OQ/PQ)
  • Microcalibrator Flowmeter
U.S. Patents 5,513,516 6,513,366 Test results may vary according to application and package specifications.